Prevention of infection with arterial catheters

Prevention of infection with arterial catheters
Replacement and relocation of catheter
Do not routinely replace peripheral arterial catheters.
Perform inspection of site daily for signs of infection, distal ischemia, or embolization.
Replace disposable or reusable transducers at 96-hours intervals. Replace other components of the system, including tubing, continuous flush device, and flush solutions at the time the transducer is changed.
Replacement of catheter-site dressing
Replace dressing when the catheter is removed or replaced, or when the dressing becomes damp, loosened, or soiled. Replace dressings more frequently in diaphoretic patients.
There is no need to remove a gauze or opaque dressing if the patient has no clinical signs of infection.
Replacement of administration sets
Replace the intravenous tubing, continuous flush device at the time the transducer is replaced (ie, 96-hour intervals).
Hang time for parenteral fluids
Replace the flush solution at the time the transducer is replaced (ie, 96-hour intervals).