Carbon dioxide monitoring (capnography)

Carbon dioxide monitoring (capnography)
Literature review current through: Jan 2024.
This topic last updated: Jan 19, 2024.

INTRODUCTION — This topic review will discuss the basic physiology and interpretation of capnography and its use in emergency settings.

DEFINITION AND BACKGROUND — The term capnography refers to the noninvasive measurement of the partial pressure of carbon dioxide (CO2) in exhaled breath expressed as the CO2 concentration over time. The relationship of CO2 concentration to time is graphically represented by the CO2 waveform, or capnogram (figure 1). Changes in the shape of the capnogram are diagnostic of disease conditions, while changes in end-tidal CO2 (EtCO2), the maximum CO2 concentration at the end of each tidal breath, can be used to assess disease severity and response to treatment. Capnography is also the most reliable indicator that an endotracheal tube is placed in the trachea after intubation.

Oxygenation and ventilation are distinct physiologic functions that must be assessed in both intubated and spontaneously breathing patients. Pulse oximetry provides instantaneous feedback about oxygenation (see "Pulse oximetry"). Capnography provides instantaneous information about ventilation (how effectively CO2 is being eliminated by the pulmonary system), perfusion (how effectively CO2 is being transported through the vascular system), and metabolism (how effectively CO2 is being produced by cellular metabolism).

Capnography became a routine part of anesthesia practice in Europe in the 1970s and in the United States in the 1980s. It is now part of the standard of care for all patients receiving general anesthesia and is part of routine monitoring in the pre-hospital and acute care settings. (see "Basic patient monitoring during anesthesia", section on 'Capnography')

PRINCIPLES OF OPERATION — Carbon dioxide (CO2) monitors measure gas concentration, or partial pressure, using one of two configurations: mainstream or sidestream. Mainstream devices measure respiratory gas (in this case CO2) directly from the airway, with the sensor located on the airway adapter at the hub of the endotracheal tube (ETT). Sidestream devices measure respiratory gas via nasal or nasal-oral cannula by aspirating a small sample from the exhaled breath through the cannula tubing to a sensor located inside the monitor (picture 1).

Mainstream systems are configured for intubated patients. Sidestream systems are configured for both intubated and nonintubated patients.

Sidestream systems are configured to use high flow rates (around 150 cc/min) or low flow rates (around 50 cc/min). Flow rates vary according to the amount of CO2 needed in the breath sample to obtain an accurate reading. Low-flow systems have a lower occlusion rate (from moisture or patient secretions) and are accurate in patients with low tidal volumes (eg, neonates, infants, and adult patients with hypoventilation and low tidal volume breathing). Low-flow systems are also resistant to dilution from supplemental oxygen. High-flow systems sampling at ≥100 cc/min have been shown to be inaccurate in neonates, infants, young children, and in hypoventilating adult patients [1-3].

CO2 monitors are either quantitative or qualitative. Quantitative devices measure the precise end-tidal CO2 (EtCO2) either as a number (capnometry) or a number and a waveform (capnography). Qualitative devices (eg, colorimetric detectors) report the range in which the EtCO2 falls (eg, 0 to 10 mmHg or >35 mmHg) as opposed to a precise value (eg, 38 mmHg).

The colorimetric EtCO2 detector, which consists of a piece of specially treated litmus paper that changes color when exposed to CO2 (purple for EtCO2 <3 mmHg; tan for 3 to 15 mmHg; and yellow for >15 mmHg). Its primary use is for verification of ETT placement. Exhalation of CO2 from an ETT placed in the trachea will change the color of the litmus paper from purple to yellow. An improperly placed ETT in the esophagus will not conduct CO2 and no change will occur in the color of the litmus paper, which will remain purple. When evaluating studies of EtCO2, it is essential to distinguish those involving qualitative measurements (colorimetric) from those describing quantitative methods with a graphic waveform display (capnography). (See "Rapid sequence intubation in adults for emergency medicine and critical care", section on 'Placement with proof'.)

Capnography uses infrared (IR) radiation to make measurements. Molecules of CO2 absorb IR radiation at a very specific wavelength (4.26 µm), with the amount of radiation absorbed having a nearly exponential relation to the CO2 concentration present in the breath sample. Detecting these changes in IR radiation levels, using appropriate photo-detectors sensitive in this spectral region, allows for the calculation of the CO2 concentration in the gas sample.

CO2 WAVEFORM — The capnogram consists of four phases (figure 1).

Phase 1 (dead space ventilation, A-B) represents the beginning of exhalation where the dead space is cleared from the upper airway and the CO2 concentration approaches zero.

Phase 2 (ascending phase, B-C) represents the rapid rise in carbon dioxide (CO2) concentration in the breath stream as the CO2 from the alveoli reaches the upper airway.

Phase 3 (alveolar plateau, C-D) represents the CO2 concentration reaching a uniform level in the entire breath stream from alveolus to nose. Point D, occurring at the end of the alveolar plateau, represents the maximum CO2 concentration at the end of the tidal breath and is appropriately named the end-tidal CO2 (EtCO2). This is the number that appears on the monitor display.

Phase 4 (D-E) represents the inspiratory cycle where the CO2 falls back to zero.

A normal capnogram (ie, a valid breath), for patients of all ages, is characterized by a set of specific elements: the CO2 concentration starts at zero and returns to zero, a maximum CO2 concentration is reached with each breath (ie, EtCO2), the amplitude is a function of the EtCO2 concentration, the width is a function of the expiratory time, and there is a characteristic shape with normal lung function.

Patients with normal lung function have characteristic trapezoidal capnograms and narrow gradients between their alveolar CO2 (ie, EtCO2) and arterial CO2 concentration (PaCO2) of 0 to 5 mmHg (figure 1). Gas in the physiologic dead space accounts for this normal gradient.

Patients with obstructive lung disease have impaired expiratory flow and uneven emptying of alveoli due to ventilation-perfusion mismatch, and demonstrate a more rounded ascending phase and an upward slope in the alveolar plateau (figure 2). In patients with abnormal lung function and ventilation-perfusion mismatch, the EtCO2-PaCO2 gradient widens depending on the severity of the lung disease [4,5]. The EtCO2 in patients with lung disease is only useful for assessing trends in ventilatory status over time; isolated EtCO2 values may or may not correlate with the PaCO2.

CLINICAL APPLICATIONS FOR INTUBATED PATIENTS — Capnography can be used in intubated patients for:

Verification of endotracheal tube (ETT) placement

Continuous monitoring of tube location during transport

Gauging effectiveness of resuscitation and prognosis during cardiac arrest

Indicator of ROSC during chest compressions

Titrating end-tidal carbon dioxide (EtCO2) levels in patients with suspected increases in intracranial pressure

Determining prognosis in trauma

Determining adequacy of ventilation

A table summarizing capnographic findings in different clinical scenarios is provided (figure 3A-B).

Verification of ETT placement — We advocate capnography in all patients who are intubated since it is highly accurate in determining ETT location and excluding unrecognized misplaced intubation (UMI). Correct ETT placement must be verified immediately following every tracheal intubation. The role of EtCO2 in confirming ETT placement is discussed below; other methods and our recommendations for verifying ETT placement are discussed in detail separately. (See "Direct laryngoscopy and endotracheal intubation in adults", section on 'Confirming proper tracheal tube placement'.)

UMI is defined as the placement of an ETT in a location other than the trachea that is unrecognized by the clinician. UMI has disastrous consequences and has been extensively documented in the emergency medical services (EMS) literature, with reported rates ranging from 7 to 25 percent [6-18]. UMI is unacceptable and widespread use of CO2 monitoring along with other verification techniques has helped to prevent it.

Following intubation, the sustained presence of a waveform with all four phases (figure 1) indicates the ETT is through the vocal cords [19]. A normal waveform can occur when the tube has been placed in the right mainstem bronchus. In addition, a waveform resembling tracheal placement may be present for a few breaths with an ETT placed in the hypopharynx just above the vocal cords, but over time these waveforms are likely to become erratic due to movement of the ETT, suggesting that the ETT is not properly placed. A four-phase capnogram may be present upon initial intubation in cardiac arrest patients, indicating that the tube is correctly positioned in the trachea [20,21].

To be considered sustained exhalation of CO2, the capnogram should meet all of the following [19]:

Amplitude rises during exhalation and falls during inspiration

Consistent or increasing amplitude over at least seven breaths

Peak amplitude more than 7.5 mmHg (1 kPa) above baseline

Reading is clinically appropriate

A flatline waveform generally indicates esophageal placement but can occur, albeit uncommonly, in several other situations, including:

ETT obstruction

Complete airway obstruction distal to the ETT (eg, foreign body)

Technical malfunction of the monitor or tubing

Prolonged cardiac arrest with diffuse cellular death (in which no CO2 is produced because of an absence of cellular respiration and CO2 exchange in the pulmonary bed is severely compromised). In a cardiac arrest patient who initially had a four-phase capnogram, tube placement should be verified if a flatline tracing subsequently develops.

The accuracy of EtCO2 to confirm tracheal location of an ETT varies with the technology used and the condition of the patient. For patients who are not in cardiac arrest, studies of qualitative colorimetric EtCO2 and quantitative capnography have demonstrated 100 percent sensitivity and 100 percent specificity for tracheal placement [22-27]. However, qualitative colorimetric EtCO2 is slightly less accurate compared with continuous quantitative waveform capnography; we prefer the latter when feasible [19]. If only qualitative colorimetric EtCO2 is used, color change must be visualized on at least five breaths since the esophagus may yield small but detectable amounts of CO2 during the first few positive-pressure ventilations.

In contrast, clinical assessment using physical signs is unreliable for determining ETT location. Fogging or condensation of the tube occurs in 83 percent of esophageal intubations [28], and chest wall movement can be produced by tracheal or esophageal tubes [29]. Anesthesiologists, under ideal operating room conditions, using breath sounds as the sole means of verification, incorrectly identify ETT location in 16 percent of cases [30].

Although the accuracy of EtCO2 for verifying ETT placement in patients with spontaneous circulation approaches 100 percent, sensitivities for tracheal placement vary greatly in cardiac arrest patients, ranging from 62 to 100 percent, depending on the modality used and the duration of the arrest [4,5,10,24,25,28-31]. In patients with low perfusion states and cadavers, studies using quantitative modalities (ie, capnography) for determining tracheal location of the ETT have shown 100 percent sensitivity [20,32-34]. Studies using qualitative (ie, colorimetric) EtCO2 methods have shown variable sensitivity, because the exhaled CO2 concentration can fall below the detection threshold [10,24,25,30]. The small number of esophageal intubations in studies assessing the accuracy of capnography for determining ETT placement in cardiac arrest patients limits the strength of these studies. Capnography's true specificity for detecting esophageal intubation in cardiac arrest patients is not certain.

When a capnogram is present in an intubated patient in cardiac arrest, the ETT can be assumed to be within the trachea [34-36]. An absent (flatline) waveform may indicate esophageal or placement above the trachea. Early in cardiac arrest, and when the patient is receiving effective CPR, complete failure of CO2 exchange is unlikely, and the clinician should assume esophageal intubation exists if a waveform is absent. In patients with prolonged cardiac arrest and an absent waveform, the clinician should also assume an esophageal intubation exists. Physical examination criteria, such as repeat laryngoscopy and auscultation of the lungs and epigastrium, may be used, but the clinician must be aware of their limitations in discriminating between esophageal and tracheal intubation. (See "Direct laryngoscopy and endotracheal intubation in adults", section on 'Confirming proper tracheal tube placement'.)

In addition to verification and monitoring of ETT placement, capnography provides similar information about the placement of alternate advanced airway devices (laryngeal tube airway, laryngeal mask airway, esophageal-tracheal combitube).

The accuracy of EtCO2 for confirming ETT location can help clinicians faced with difficult airway management decisions. As an example, in a case involving one of the authors, a patient resuscitated following a submersion injury became difficult to ventilate and hypotensive following RSI, but the presence of a waveform reassured clinicians about ETT placement, allowing them to consider other reasons for decompensation. Decompression of the patient’s distended stomach with an orogastric tube resolved the difficulties with ventilation and blood pressure.

Monitoring ETT location during transport — UMI has catastrophic consequences and can occur when an ETT is dislodged during patient transport. Continuous monitoring of ETT location during transport prevents UMI. EtCO2 confirmation of initial ETT placement and continuous capnographic monitoring of ETT location is an accepted standard of care by the American Society of Anesthesiologists [37] and is recommended as the most reliable method by other national organizations [38-40].

A prospective, observational study of 153 prehospital intubations found a 23 percent UMI rate among patients without continuous EtCO2 monitoring, while patients with continuously monitored EtCO2 sustained the desired 0 percent UMI rate [18]. The attached figures depict one airway confirmation algorithm using colorimetric CO2 detectors and another using capnography. They can be used for both prehospital and in-hospital intubated patients.

Effectiveness of CPR — In the 1980s, studies using animal models showed that EtCO2 levels reflect cardiac output during cardiopulmonary resuscitation (CPR) and can be used as a noninvasive measure of cardiac output. A landmark study in 1988 demonstrated this principle in humans [41]. During cardiac arrest, when alveolar ventilation and metabolism are essentially constant, EtCO2 reflects pulmonary blood flow. Therefore, EtCO2 can be used as a gauge of the effectiveness of cardiac compressions. As effective CPR leads to a higher cardiac output, EtCO2 will rise, reflecting the increase in perfusion.

The measurement of EtCO2 varies directly with the cardiac output produced by chest compression and has been described in both EMS [42] and ICU patients [41]. Both of these prospective, observational studies found an EtCO2 level <3 mmHg immediately after cardiac arrest, with a higher level generated during cardiac compressions and a mean peak >7.5 mmHg just before return of spontaneous circulation (ROSC) [41,42]. This peak in EtCO2 level is the earliest sign of ROSC and may occur before return of a palpable pulse or blood pressure. Another observational study of data collected during human CPR found a positive correlation between the depth of chest compressions and EtCO2 (10 mm increase in compression depth increased EtCO2 by 1.4 mmHg), as well as ventilation rate and EtCO2 (increase of 10 breaths/minute decreased ETCO2 by 3 mmHg), and confirmed that higher EtCO2 values during CPR correlate with increased ROSC and survival [43].

Return of spontaneous circulation — During cardiac arrest, EtCO2 is the earliest indicator of the return of spontaneous circulation (ROSC) [41,42]. When the heart restarts, the dramatic increase in cardiac output, and resulting increase in perfusion, leads to a rapid increase in EtCO2 as the CO2 that has accumulated during cardiac arrest is effectively transported to the lungs and exhaled. This process manifests as a sudden rise in EtCO2. (See 'Effectiveness of CPR' above.)

American Heart Association (AHA) guidelines for cardiac resuscitation emphasize the importance of continuing chest compressions without interruption until a perfusing rhythm is reestablished. Experimental evidence indicates that interruptions in chest compressions are followed by sustained periods of reduced blood flow, which only gradually return to pre-interruption levels. Capnogram monitoring virtually eliminates the need to stop chest compressions to check for pulses. Reestablishment of a perfusing rhythm is accompanied immediately by a dramatic increase in EtCO2. Once this rise in EtCO2 is noted, chest compressions can safely be stopped while cardiac rhythm and blood pressure are assessed [40,44].

According to an observational study of 145 patients with out of hospital cardiac arrest (OHCA), several seconds may be required after electrical conversion to a potentially perfusing rhythm before effective mechanical contractions and a subsequent rise in EtCO2 occur [45]. Accordingly, stopping compressions in response to a rhythm change in order to check for a pulse may result in degradation of the rhythm, as coronary perfusion had not yet been restored. EtCO2 is a more reliable harbinger of reperfusion than a pulse check. (See "Adult basic life support (BLS) for health care providers".)

Specific EtCO2 levels have been shown to correlate with ROSC. In a systematic review of 17 observational studies including over 6100 adults with cardiac arrest, which included metaanalyses of data from five studies, EtCO2 ≥10 to 20 mmHg during CPR was strongly associated with ROSC while persistent EtCO2 below 10 mmHg after 20 minutes of CPR had a 0.5 percent likelihood of ROSC [46]. Subsequent retrospective studies support this general finding [47,48]. A retrospective, observational study of 526 OHCA patients with a nonshockable rhythm found that the first ETCO2 measured after placement of an airway device predicted ROSC and survival after hospital admission [47]. Specifically, patients with an ETCO2 >45 mmHg had an increased probability of sustained ROSC and 30-day survival. A retrospective study of 324 OHCA patients found that higher EtCO2 levels were associated with higher rates of successful defibrillation [48].

Confounding factors in resuscitation — Drugs used in resuscitation may affect EtCO2 values. EtCO2 often decreases rapidly moments after administration of epinephrine. Sodium bicarbonate may produce a transient increase in EtCO2, but the rise in EtCO2 levels after ROSC is greater and longer lasting than after a sodium bicarbonate bolus [49,50].

Prognosis in cardiac arrest — In several prospective, observational studies, EtCO2 levels of ≤10 mmHg measured 20 minutes after the initiation of advanced cardiac life support accurately predicted death in adult patients with cardiac arrest [51-54]. The prognostic value of measuring EtCO2 has been demonstrated in animal [55] and human studies [51-54,56].

Cause of cardiac arrest — EtCO2 may also be helpful in determining the etiology of cardiac arrest. Two animal studies reported higher EtCO2 levels at the onset of cardiac arrest caused by primary asphyxia than arrest caused by ventricular fibrillation [57,58]. Researchers found results consistent with these animal experiments in a prospective, observational study of prehospital victims of cardiac arrest. Patients in the asphyxia group (initial rhythm of asystole or pulseless electrical activity associated with conditions such as airway foreign body, aspiration, asthma, or drowning) had higher EtCO2 levels compared with patients in the ventricular tachycardia/fibrillation group (initial rhythm of ventricular tachycardia/fibrillation associated with acute myocardial infarction) [59]. These between group differences disappear within minutes if return of spontaneous circulation has not occurred.

Increased ICP and trauma prognosis — EtCO2 monitoring can help clinicians avoid inadvertent hyperventilation of patients with traumatic brain injury and suspected increased intracranial pressure (ICP). It may also help determine the prognosis of trauma victims.

Arterial CO2 tension affects blood flow to the brain. High CO2 levels result in cerebral vasodilation, while low CO2 levels result in cerebral vasoconstriction. Sustained hypoventilation (defined as PaCO2 levels ≥50 mmHg) results in increased cerebral blood flow and increased ICP, which can harm head-injured patients. Sustained hyperventilation (defined as PaCO2 ≤30 mmHg) is also detrimental and is associated with worse neurologic outcome in severely brain-injured patients. Consequently, ventilation rates to achieve eucapnia are recommended by the brain trauma foundation [60].

Use of continuous EtCO2 monitoring in intubated trauma patients is supported by several studies:

A prospective observational study found a lower incidence of inadvertent hyperventilation among intubated patients with severe head injury who underwent continuous EtCO2 monitoring compared with those without EtCO2 monitoring (5.6 versus 13.4 percent; odds ratio [OR] 2.64; 95% CI 1.12-6.20) [61].

In a controlled study of intubated blunt trauma victims, patients assigned to capnography-guided ventilation during transport were significantly more likely to arrive at the emergency department appropriately ventilated, based on PaCO2 levels obtained by arterial blood gas [62].

Studies have found that a low EtCO2 in trauma patients is associated with mortality and hemorrhagic shock:

In a prospective observational study of blunt trauma patients requiring prehospital intubation, EtCO2 values obtained 20 minutes after intubation distinguished the great majority of survivors from non-survivors [63]. Median EtCO2 among survivors was 30.8 mmHg and among non-survivors 26.3 mmHg (95 percent CI of difference between medians 3 to 6.75 mmHg).

In a retrospective observational study of 135 trauma patients transported by emergency medical services to a level one trauma center, the mean prehospital EtCO2 level distinguished survivors (34 mmHg, 95% CI 32-35) from non-survivors (18 mmHg, 95% CI 9-28) and had better discrimination regarding prognosis than vital signs or pulse oximetry [64].

In a retrospective review of 373 trauma patients receiving a massive transfusion protocol and undergoing damage control surgery, operative ETCO2 ≤23mmHg was a predictor of mortality [65].

In two retrospective studies of non-intubated adult trauma patients, initial side-stream ETCO2 ≤28.5 to 29.5 mmHg was associated with higher in-hospital mortality, blood transfusion requirement, and complication rates [66,67].

CLINICAL APPLICATIONS FOR SPONTANEOUSLY BREATHING PATIENTS — In a spontaneously breathing, nonintubated patient, capnography can be used for the following:

Performing rapid assessment of critically ill or seizing patients

Determining response to treatment in acute respiratory distress

Determining adequacy of ventilation in obtunded or unconscious patients, or in patients undergoing procedural sedation

Detecting metabolic acidosis in diabetic patients and in children with gastroenteritis

Providing prognostic indicators in patients with sepsis or septic shock

A table summarizing capnographic findings in different clinical scenarios is provided (figure 3A-B).

Assessing airway, breathing, and circulation — The airway, breathing, and circulation (ABCs) of patients can be rapidly assessed using the capnography waveform and end-tidal carbon dioxide (EtCO2) values [68]. The presence of a normal waveform denotes a patent airway and spontaneous breathing [69]. Normal EtCO2 levels (35 to 45 mmHg) signify adequate ventilation and perfusion [41,70]. Capnography can be used to assess unresponsive patients ranging from those who are actively seizing to victims of chemical terrorism (table 1) [71]. Unlike pulse oximetry, capnography does not misinterpret motion artifact and provides reliable readings in low perfusion states. Studies have found that an EtCO2 ≤29.5 mmHg was independently predictive of the need for massive transfusion [66,72].

Seizure — Capnography is the only monitoring modality that is accurate and reliable in actively seizing patients because the capnogram is determined entirely by respiratory activity and is not confounded by muscle activity or movement artifact. Capnographic data (respiratory rate, EtCO2, and capnogram) can be used to distinguish among:

Seizing patients with apnea (flatline waveform, no EtCO2 reading, and no chest wall movement)

Seizing patients with ineffective ventilation (small waveforms, low EtCO2 values), and

Seizing patients with effective ventilation (normal CO2 waveform, normal EtCO2 values). (See "Convulsive status epilepticus in adults: Classification, clinical features, and diagnosis" and "Convulsive status epilepticus in adults: Management".)

Acute respiratory distress — Capnography provides dynamic monitoring of ventilatory status in patients with acute respiratory distress from any cause, including: bronchiolitis, asthma, cystic fibrosis, heart failure, and chronic obstructive pulmonary disease (COPD). Several studies have found that quantitative waveform analysis can discriminate between heart failure and COPD [73-75].

By measuring EtCO2 and respiratory rate with each breath, capnography provides instantaneous feedback on the clinical status of the patient. Respiratory rate is measured directly from the airway (nose and mouth) with oral or nasal cannula, providing a more reliable reading than impedance respiratory monitoring. In upper airway obstruction and laryngospasm, impedance monitoring detects chest wall movement, interprets this as a valid breath, and displays a respiratory rate, even though the patient is not ventilating. In contrast, capnography detects no ventilation and shows a flatline waveform.

Clinicians can rapidly assess EtCO2 trends. A patient with a respiratory rate of 30 will generate 150 EtCO2 readings in five minutes. This provides sufficient information to determine whether the patient's ventilation is worsening despite treatment (increasing EtCO2), stabilizing (stable EtCO2), or improving (decreasing EtCO2). As an example, increasing EtCO2 would reflect the worsening ventilation of an acutely tachypneic patient with obstructive lung disease who is developing respiratory muscle fatigue or a more severe lower airway obstruction.

Procedural sedation — We recommend continuous capnographic monitoring in all patients undergoing procedural sedation and analgesia (PSA). Capnography can rapidly detect the common adverse airway and respiratory events associated with PSA, including respiratory depression, apnea, upper airway obstruction, laryngospasm, and bronchospasm. Patients monitored with capnography experience significantly fewer episodes of hypoxia than those monitored in standard fashion, especially since the administration of supplemental oxygen renders pulse oximetry less effective as an early warning device for respiratory depression during PSA. (See "Procedural sedation in adults in the emergency department: General considerations, preparation, monitoring, and mitigating complications", section on 'Monitoring' and "Procedural sedation in children: Approach", section on 'Monitoring'.)

Respiratory depression caused by oversedation will manifest an abnormally high or low EtCO2 before pulse oximetry detects a falling oxyhemoglobin saturation, especially in patients receiving supplemental oxygen. In one trial, the median time between the onset of respiratory depression, as determined by EtCO2, and hypoxia was 60 seconds (range 5 to 240 seconds) [76].

Common capnographic findings during PSA include the following and are summarized in the figures (figure 3A and figure 3B):

Flatline waveform – This can occur from monitor calibration (will indicate "calibrating" on the monitor), cannula occlusion (will indicate "occlusion" on the monitor), or apnea (central or obstructive). The combination of absent chest wall movement and a flatline waveform differentiates central apnea from obstructive apnea (upper airway obstruction or laryngospasm), which manifests chest wall movement. Response to airway alignment maneuvers (eg, chin lift, jaw thrust) can distinguish upper airway obstruction from laryngospasm.

Oscillating waveform progressively decreasing in height – This occurs with oversedation and is a sign of impending apnea.

Small waveforms interspersed with flatline waveforms - This irregular waveform pattern occurs with hypopneic hypoventilation and indicates periodic breathing.

Increasing waveform height – This occurs with bradypneic hypoventilation and reflects increasing EtCO2. An EtCO2 >70 mmHg in a patient without COPD indicates respiratory failure.

Decreasing waveform height – This occurs with hypopneic hypoventilation (ie, low tidal volume breathing) and reflects decreasing EtCO2.

Increasing waveform width – This occurs with hypoventilation and increasing expiratory time.

Decreasing waveform width – This occurs with hyperventilation and decreasing expiratory time.

Waveform does not return to zero – An off-baseline waveform can occur with rebreathing.

Curved/shark-fin appearing waveform – The characteristic curved capnogram (figure 2) can be seen in a patient with obstructive lung disease (eg, asthma, COPD, cystic fibrosis) and indicates bronchospasm if it occurs in a patient without chronic lung disease.

Ventilation in obtunded or unconscious patients — Patients with altered mental status, including those with alcohol intoxication, intentional or unintentional drug overdose, and postictal patients (especially those treated with benzodiazepines), may have impaired ventilatory function. Capnography can differentiate between patients with effective ventilation and those with ineffective ventilation [77].  

Detecting metabolic acidosis — There is a positive linear correlation between serum bicarbonate (HCO3) and EtCO2 in patients with diabetic ketoacidosis and children with gastroenteritis [78-82]. As the patient becomes acidotic, HCO3 decreases, causing an increase in minute ventilation, which results in a compensatory respiratory alkalosis. This process results in a decrease in EtCO2.

By increasing minute ventilation, these patients are able to lower arterial CO2 tension to help correct their underlying acidemia. The more acidotic the patient, the lower the HCO3, the higher the respiratory rate, and the lower the EtCO2. Capnography can be used as an indicator of metabolic acidosis in these patients. In addition, EtCO2 can be used to distinguish diabetics in ketoacidosis (metabolic acidosis, compensatory tachypnea, and low EtCO2) from those who are not (nonacidotic, normal respiratory rate, and normal EtCO2) [83,84]. (See "Approach to the adult with metabolic acidosis" and "Approach to the child with metabolic acidosis".)

Prognosis in sepsis — There is an inverse relationship between EtCO2 and lactate levels in sepsis, severe sepsis, and septic shock. EtCO2 performs similarly to lactate as a predictor for mortality in patients with suspected sepsis [85,86]. In a prospective study of 183 patients performed in the prehospital setting, patients with suspected sepsis and an EtCO2 value ≤25 mmHg were more frequently diagnosed with sepsis (78 versus 43 percent) and severe sepsis (47 versus 7 percent), and had higher mortality (11 versus 5 percent) [87]. The use of EtCO2 to predict mortality, sepsis, and severe sepsis in prehospital patients outperformed another well-accepted screening tool, the quick sequential organ failure assessment score (qSOFA) [88]. Patients with severely compromised respiratory reserve who cannot appropriately compensate for the lactic acidosis resulting from severe sepsis or septic shock have the worst prognosis despite ETCO2 values that are not low. (See "Evaluation and management of suspected sepsis and septic shock in adults".)

LIMITATIONS — Capnography is most effective when assessing a pure ventilation, perfusion, or metabolism problem. Capnographic findings in mixed ventilation, perfusion, or metabolism problems are difficult to interpret. In patients with complex pathophysiology, a ventilation problem may elevate end-tidal carbon dioxide (EtCO2), while a perfusion problem may simultaneously lower EtCO2. As an example, capnography may not provide useful information in a patient with acute pulmonary edema (ventilation problem) from an underlying acute myocardial infarction with decreased cardiac output (perfusion problem).

In high-flow CO2 systems, when the tidal volume of the patient drops below the flow rate of the system, the monitor will entrain room air to compensate, falsely diluting the EtCO2 reading and slurring the ascending phase of the waveform. For this and other reasons, low flow systems are preferred. (See 'Principles of operation' above.)

Capnography is highly accurate for determining tracheal placement of an endotracheal tube (ETT) in nearly all clinical circumstances. However, the test characteristics of capnography in cardiac arrest can vary, and this is discussed above. (See 'Verification of ETT placement' above.)

SUMMARY AND RECOMMENDATIONS

Definition and background – Carbon dioxide (CO2) monitoring is a versatile noninvasive diagnostic tool for continuous assessment of the ventilatory status of both intubated and nonintubated patients. Capnography provides instantaneous information about ventilation (how effectively CO2 is being eliminated by the pulmonary system), perfusion (how effectively CO2 is being transported through the vascular system), and metabolism (how effectively CO2 is being produced by cellular metabolism). A table summarizing capnographic findings in different clinical scenarios is provided (figure 3A-B). (See 'Definition and background' above.)

Principles of operation – Colorimetric detectors provide a qualitative measurement of end-tidal CO2 (EtCO2); capnometry provides a quantitative EtCO2 measurement; capnography provides a quantitative measure and a waveform depicting CO2 levels over time. (See 'Principles of operation' above.)

CO2 waveform – The capnogram consists of four phases: dead space ventilation, ascending phase, alveolar plateau, and descending inspiratory phase, which are depicted in the figure (figure 1). (See 'CO2 Waveform' above.)

Clinical applications – Capnography is of greatest value for emergency departments and emergency medical services for confirming placement of endotracheal tubes (ETTs), continuous monitoring of tube position during transport, assessing the status of patients in cardiac arrest, and monitoring the respiratory status of patients undergoing procedural sedation. (See 'Clinical applications for intubated patients' above.)

Verification of ETT placement – In all intubated patients, we recommend that an EtCO2 monitor be used to help confirm proper placement of the ETT. EtCO2 monitoring is the primary means of confirming tracheal placement of the ETT. Clinical techniques may be used adjunctively, but all have significant limitations in discriminating between esophageal and tracheal intubation. (See 'Verification of ETT placement' above.)

Following intubation, the presence of a sustained waveform with all four phases (figure 1) is a highly accurate indication that the ETT is placed in the trachea, including upon initial intubation in cardiac arrest patients. If only qualitative colorimetric EtCO2 is used, color change must be visualized on at least five breaths since the esophagus may yield small but detectable amounts of CO2 during the first few positive-pressure ventilations.

A flatline waveform generally indicates esophageal placement but can occur with: ETT obstruction, complete airway obstruction distal to the ETT (eg, foreign body), technical malfunction of the monitor or tubing, and prolonged cardiac arrest. When the amplitude of the tracing is very small or CO2 is not detectable by waveform or qualitative EtCO2 detection, clinicians should assume an esophageal intubation exists.

Monitoring resuscitation in cardiac arrest – In a patient in cardiac arrest, we suggest that EtCO2 be used to monitor resuscitation efforts. EtCO2 monitoring can be performed on patients in cardiac arrest to determine the adequacy of chest compressions and the return of spontaneous circulation. During cardiac arrest, EtCO2 reflects pulmonary blood flow and can be used as a gauge of the effectiveness of cardiac compressions. As effective CPR leads to a higher cardiac output, EtCO2 will rise. (See 'Effectiveness of CPR' above.)

Assessing return of spontaneous circulation – EtCO2 is also the earliest indicator of the return of spontaneous circulation. When the heart restarts, the dramatic increase in cardiac output, and resulting increase in perfusion, leads to a rapid increase in EtCO2. (See 'Return of spontaneous circulation' above.)

Monitoring during procedural sedation – We recommend continuous capnographic monitoring in all patients undergoing procedural sedation and analgesia (PSA). Capnography can rapidly detect common adverse airway and respiratory events including central apnea, obstructive apnea (upper airway obstruction, laryngospasm), bronchospasm, and respiratory failure (algorithm 1). The evidence for capnographic monitoring is presented separately. (See 'Procedural sedation' above and "Procedural sedation in adults in the emergency department: General considerations, preparation, monitoring, and mitigating complications", section on 'Monitoring'.)

Applications in spontaneously breathing patients – Capnography can be helpful in monitoring the status of patients with seizures, acute respiratory distress, unconsciousness, and metabolic acidosis. (See 'Clinical applications for spontaneously breathing patients' above.)

Limitations – Capnography is most effective when assessing a pure ventilation, perfusion, or metabolism problem. Capnographic findings in mixed ventilation, perfusion, or metabolism problems are difficult to interpret. (See 'Limitations' above.)

ACKNOWLEDGMENT — The editorial staff at UpToDate acknowledge Salvatore Silvestri, MD, now deceased, who contributed to an earlier version of this topic review.

  1. Friesen RH, Alswang M. End-tidal PCO2 monitoring via nasal cannulae in pediatric patients: accuracy and sources of error. J Clin Monit 1996; 12:155.
  2. Gravenstein N. Capnometry in infants should not be done at lower sampling flow rates. J Clin Monit 1989; 5:63.
  3. Sasse FJ. Can we trust end-tidal carbon dioxide measurements in infants? J Clin Monit 1985; 1:147.
  4. Yamanaka MK, Sue DY. Comparison of arterial-end-tidal PCO2 difference and dead space/tidal volume ratio in respiratory failure. Chest 1987; 92:832.
  5. Hardman JG, Aitkenhead AR. Estimating alveolar dead space from the arterial to end-tidal CO(2) gradient: a modeling analysis. Anesth Analg 2003; 97:1846.
  6. Stewart RD, Paris PM, Winter PM, et al. Field endotracheal intubation by paramedical personnel. Success rates and complications. Chest 1984; 85:341.
  7. Shea SR, MacDonald JR, Gruzinski G. Prehospital endotracheal tube airway or esophageal gastric tube airway: a critical comparison. Ann Emerg Med 1985; 14:102.
  8. Pointer JE. Clinical characteristics of paramedics' performance of endotracheal intubation. J Emerg Med 1988; 6:505.
  9. Jenkins WA, Verdile VP, Paris PM. The syringe aspiration technique to verify endotracheal tube position. Am J Emerg Med 1994; 12:413.
  10. Bozeman WP, Hexter D, Liang HK, Kelen GD. Esophageal detector device versus detection of end-tidal carbon dioxide level in emergency intubation. Ann Emerg Med 1996; 27:595.
  11. Sayre MR, Sakles JC, Mistler AF, et al. Field trial of endotracheal intubation by basic EMTs. Ann Emerg Med 1998; 31:228.
  12. Eckstein M, Chan L, Schneir A, Palmer R. Effect of prehospital advanced life support on outcomes of major trauma patients. J Trauma 2000; 48:643.
  13. Wang HE, Sweeney TA, O'Connor RE, Rubinstein H. Failed prehospital intubations: an analysis of emergency department courses and outcomes. Prehosp Emerg Care 2001; 5:134.
  14. Katz SH, Falk JL. Misplaced endotracheal tubes by paramedics in an urban emergency medical services system. Ann Emerg Med 2001; 37:32.
  15. Gausche M, Lewis RJ, Stratton SJ, et al. Effect of out-of-hospital pediatric endotracheal intubation on survival and neurological outcome: a controlled clinical trial. JAMA 2000; 283:783.
  16. Jones JH, Murphy MP, Dickson RL, et al. Emergency physician-verified out-of-hospital intubation: miss rates by paramedics. Acad Emerg Med 2004; 11:707.
  17. Jemmett ME, Kendal KM, Fourre MW, Burton JH. Unrecognized misplacement of endotracheal tubes in a mixed urban to rural emergency medical services setting. Acad Emerg Med 2003; 10:961.
  18. Silvestri S, Ralls GA, Krauss B, et al. The effectiveness of out-of-hospital use of continuous end-tidal carbon dioxide monitoring on the rate of unrecognized misplaced intubation within a regional emergency medical services system. Ann Emerg Med 2005; 45:497.
  19. Chrimes N, Higgs A, Hagberg CA, et al. Preventing unrecognised oesophageal intubation: a consensus guideline from the Project for Universal Management of Airways and international airway societies. Anaesthesia 2022; 77:1395.
  20. Silvestri S, Ladde JG, Brown JF, et al. Endotracheal tube placement confirmation: 100% sensitivity and specificity with sustained four-phase capnographic waveforms in a cadaveric experimental model. Resuscitation 2017; 115:192.
  21. Reid C, Ferguson I, Burns B, et al. Does end-tidal capnography confirm tracheal intubation in fresh-frozen cadavers? Emerg Med J 2017; 34:315.
  22. Goldberg JS, Rawle PR, Zehnder JL, Sladen RN. Colorimetric end-tidal carbon dioxide monitoring for tracheal intubation. Anesth Analg 1990; 70:191.
  23. Knapp S, Kofler J, Stoiser B, et al. The assessment of four different methods to verify tracheal tube placement in the critical care setting. Anesth Analg 1999; 88:766.
  24. Ornato JP, Shipley JB, Racht EM, et al. Multicenter study of a portable, hand-size, colorimetric end-tidal carbon dioxide detection device. Ann Emerg Med 1992; 21:518.
  25. MacLeod BA, Heller MB, Gerard J, et al. Verification of endotracheal tube placement with colorimetric end-tidal CO2 detection. Ann Emerg Med 1991; 20:267.
  26. Vukmir RB, Heller MB, Stein KL. Confirmation of endotracheal tube placement: a miniaturized infrared qualitative CO2 detector. Ann Emerg Med 1991; 20:726.
  27. Grmec S, Mally S. Prehospital determination of tracheal tube placement in severe head injury. Emerg Med J 2004; 21:518.
  28. Kelly JJ, Eynon CA, Kaplan JL, et al. Use of tube condensation as an indicator of endotracheal tube placement. Ann Emerg Med 1998; 31:575.
  29. Pollard BJ, Junius F. Accidental intubation of the oesophagus. Anaesth Intensive Care 1980; 8:183.
  30. Birmingham PK, Cheney FW, Ward RJ. Esophageal intubation: a review of detection techniques. Anesth Analg 1986; 65:886.
  31. Bhende MS, Thompson AE. Evaluation of an end-tidal CO2 detector during pediatric cardiopulmonary resuscitation. Pediatrics 1995; 95:395.
  32. Sayah AJ, Peacock WF, Overton DT. End-tidal CO2 measurement in the detection of esophageal intubation during cardiac arrest. Ann Emerg Med 1990; 19:857.
  33. Silvestri S, Krauss B, Ralls G, et al. Emergency department capnographic confirmation of prehospital endotracheal intubation in cardiac arrest patients - A preliminary report. Ann Emerg Med 2005; 46:6.
  34. Grmec S. Comparison of three different methods to confirm tracheal tube placement in emergency intubation. Intensive Care Med 2002; 28:701.
  35. Reid C, Lewis A, Habig K, et al. Sustained life-like waveform capnography after human cadaveric tracheal intubation. Emerg Med J 2015; 32:232.
  36. Cook TM, Nolan JP. Use of capnography to confirm correct tracheal intubation during cardiac arrest. Anaesthesia 2011; 66:1183.
  37. American Society of Anesthesiologists. Basic Standards for Preanesthesia Care 1999. American Society of Anesthesiologists: Standards for basic anesthetic monitoring. http://www.asahq.org/publicationsAndServices/standards/32.html (Accessed on April 16, 2006).
  38. O'Connor RE, Swor RA. Verification of endotracheal tube placement following intubation. National Association of EMS Physicians Standards and Clinical Practice Committee. Prehosp Emerg Care 1999; 3:248.
  39. American College of Emergency Physicians Policy Statement. Verification of Endotracheal Tube Placement. http://www.acep.org/webportal/PracticeResources/PolicyStatements/pracmgt/VerificationofEndotrachealTubePlacement.htm (Accessed on April 16, 2006).
  40. Neumar RW, Otto CW, Link MS, et al. Part 8: adult advanced cardiovascular life support: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation 2010; 122:S729.
  41. Falk JL, Rackow EC, Weil MH. End-tidal carbon dioxide concentration during cardiopulmonary resuscitation. N Engl J Med 1988; 318:607.
  42. Garnett AR, Ornato JP, Gonzalez ER, Johnson EB. End-tidal carbon dioxide monitoring during cardiopulmonary resuscitation. JAMA 1987; 257:512.
  43. Sheak KR, Wiebe DJ, Leary M, et al. Quantitative relationship between end-tidal carbon dioxide and CPR quality during both in-hospital and out-of-hospital cardiac arrest. Resuscitation 2015; 89:149.
  44. 2005 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care, Part 7. 1: Adjuncts for airway control and ventilation. Circulation 2005; 112(suppl IV):IV.
  45. Davis DP, Sell RE, Wilkes N, et al. Electrical and mechanical recovery of cardiac function following out-of-hospital cardiac arrest. Resuscitation 2013; 84:25.
  46. Paiva EF, Paxton JH, O'Neil BJ. The use of end-tidal carbon dioxide (ETCO2) measurement to guide management of cardiac arrest: A systematic review. Resuscitation 2018; 123:1.
  47. Poppe M, Stratil P, Clodi C, et al. Initial end-tidal carbon dioxide as a predictive factor for return of spontaneous circulation in nonshockable out-of-hospital cardiac arrest patients: A retrospective observational study. Eur J Anaesthesiol 2019; 36:524.
  48. Hubble MW, Van Vleet L, Taylor S, et al. Predictive Utility of End-Tidal Carbon Dioxide on Defibrillation Success in Out-of-Hospital Cardiac Arrest. Prehosp Emerg Care 2021; 25:697.
  49. Sandroni C, De Santis P, D'Arrigo S. Capnography during cardiac arrest. Resuscitation 2018; 132:73.
  50. Roh YI, Kim HI, Kim SJ, et al. End-tidal carbon dioxide after sodium bicarbonate infusion during mechanical ventilation or ongoing cardiopulmonary resuscitation. Am J Emerg Med 2023; 76:211.
  51. Sanders AB, Kern KB, Otto CW, et al. End-tidal carbon dioxide monitoring during cardiopulmonary resuscitation. A prognostic indicator for survival. JAMA 1989; 262:1347.
  52. Levine RL, Wayne MA, Miller CC. End-tidal carbon dioxide and outcome of out-of-hospital cardiac arrest. N Engl J Med 1997; 337:301.
  53. Wayne MA, Levine RL, Miller CC. Use of end-tidal carbon dioxide to predict outcome in prehospital cardiac arrest. Ann Emerg Med 1995; 25:762.
  54. Grmec S, Klemen P. Does the end-tidal carbon dioxide (EtCO2) concentration have prognostic value during out-of-hospital cardiac arrest? Eur J Emerg Med 2001; 8:263.
  55. Sanders AB, Ewy GA, Bragg S, et al. Expired PCO2 as a prognostic indicator of successful resuscitation from cardiac arrest. Ann Emerg Med 1985; 14:948.
  56. Asplin BR, White RD. Prognostic value of end-tidal carbon dioxide pressures during out-of-hospital cardiac arrest. Ann Emerg Med 1995; 25:756.
  57. Bhende MS, Karasic DG, Karasic RB. End-tidal carbon dioxide changes during cardiopulmonary resuscitation after experimental asphyxial cardiac arrest. Am J Emerg Med 1996; 14:349.
  58. Berg RA, Henry C, Otto CW, et al. Initial end-tidal CO2 is markedly elevated during cardiopulmonary resuscitation after asphyxial cardiac arrest. Pediatr Emerg Care 1996; 12:245.
  59. Grmec S, Lah K, Tusek-Bunc K. Difference in end-tidal CO2 between asphyxia cardiac arrest and ventricular fibrillation/pulseless ventricular tachycardia cardiac arrest in the prehospital setting. Crit Care 2003; 7:R139.
  60. Brain Trauma Foundation. Management and prognosis of severe traumatic brain injury guidelines. http://www2.braintrauma.org/guidelines/downloads/btf_guidelines_management.pdf (Accessed on April 29, 2006).
  61. Davis DP, Dunford JV, Ochs M, et al. The use of quantitative end-tidal capnometry to avoid inadvertent severe hyperventilation in patients with head injury after paramedic rapid sequence intubation. J Trauma 2004; 56:808.
  62. Helm M, Schuster R, Hauke J, Lampl L. Tight control of prehospital ventilation by capnography in major trauma victims. Br J Anaesth 2003; 90:327.
  63. Deakin CD, Sado DM, Coats TJ, Davies G. Prehospital end-tidal carbon dioxide concentration and outcome in major trauma. J Trauma 2004; 57:65.
  64. Childress K, Arnold K, Hunter C, et al. Prehospital End-tidal Carbon Dioxide Predicts Mortality in Trauma Patients. Prehosp Emerg Care 2018; 22:170.
  65. Demaree C, Simpson JT, Smith A, et al. Intraoperative End-Tidal CO₂ as a Predictor of Mortality in Trauma Patients Receiving Massive Transfusion Protocol. Am Surg 2019; 85:e617.
  66. Bryant MK, Portelli Tremont JN, Patel Z, et al. "Low initial pre-hospital end-tidal carbon dioxide predicts inferior clinical outcomes in trauma patients". Injury 2021; 52:2502.
  67. Portelli Tremont JN, Caldas RA, Cook N, et al. Low initial in-hospital end-tidal carbon dioxide predicts poor patient outcomes and is a useful trauma bay adjunct. Am J Emerg Med 2022; 56:45.
  68. Miller RD. Miller's Anesthesia, 6th, Elsevier Churchill Livingstone, New York 2005.
  69. Swedlow, DB. Capnometry and capnography: The anesthesia disaster early warning system. Seminars in Anesthesia 1986; 3:194.
  70. Weil MH, Bisera J, Trevino RP, Rackow EC. Cardiac output and end-tidal carbon dioxide. Crit Care Med 1985; 13:907.
  71. Krauss B. Capnography as a rapid assessment and triage tool for chemical terrorism. Pediatr Emerg Care 2005; 21:493.
  72. Stone ME Jr, Kalata S, Liveris A, et al. End-tidal CO2 on admission is associated with hemorrhagic shock and predicts the need for massive transfusion as defined by the critical administration threshold: A pilot study. Injury 2017; 48:51.
  73. Mieloszyk RJ, Verghese GC, Deitch K, et al. Automated quantitative analysis of capnogram shape for COPD-normal and COPD-CHF classification. IEEE Trans Biomed Eng 2014; 61:2882.
  74. Abid A, Mieloszyk RJ, Verghese GC, et al. Model-Based Estimation of Respiratory Parameters from Capnography, With Application to Diagnosing Obstructive Lung Disease. IEEE Trans Biomed Eng 2017; 64:2957.
  75. Koyama T, Kobayashi M, Ichikawa T, et al. Technology Applications of Capnography Waveform Analytics for Evaluation of Heart Failure Severity. J Cardiovasc Transl Res 2020; 13:1044.
  76. Deitch K, Miner J, Chudnofsky CR, et al. Does end tidal CO2 monitoring during emergency department procedural sedation and analgesia with propofol decrease the incidence of hypoxic events? A randomized, controlled trial. Ann Emerg Med 2010; 55:258.
  77. Millane T, Greene S, Rotella JA, Leang YH. End-tidal capnography provides reliable ventilatory monitoring for non-intubated patients presenting after sedative overdose to the emergency department. Emerg Med Australas 2020; 32:164.
  78. Fearon DM, Steele DW. End-tidal carbon dioxide predicts the presence and severity of acidosis in children with diabetes. Acad Emerg Med 2002; 9:1373.
  79. Garcia E, Abramo TJ, Okada P, et al. Capnometry for noninvasive continuous monitoring of metabolic status in pediatric diabetic ketoacidosis. Crit Care Med 2003; 31:2539.
  80. Nagler J, Wright RO, Krauss B. End-tidal carbon dioxide as a measure of acidosis among children with gastroenteritis. Pediatrics 2006; 118:260.
  81. Bou Chebl R, Madden B, Belsky J, et al. Diagnostic value of end tidal capnography in patients with hyperglycemia in the emergency department. BMC Emerg Med 2016; 16:7.
  82. Soleimanpour H, Taghizadieh A, Niafar M, et al. Predictive value of capnography for suspected diabetic ketoacidosis in the emergency department. West J Emerg Med 2013; 14:590.
  83. Kartal M, Eray O, Rinnert S, et al. ETCO₂: a predictive tool for excluding metabolic disturbances in nonintubated patients. Am J Emerg Med 2011; 29:65.
  84. Solmeinpur H, et al. Predictive Value of Capnography for Diagnosis in Patients with Suspected Diabetic Ketoacidosis in the Emergency Department. West J Emerg Med 2013; (www.escholarship.org/uc/item/5qz744fv).
  85. Hunter CL, Silvestri S, Dean M, et al. End-tidal carbon dioxide is associated with mortality and lactate in patients with suspected sepsis. Am J Emerg Med 2013; 31:64.
  86. Cully M, Treut M, Thompson AD, DePiero AD. Exhaled end-tidal carbon dioxide as a predictor of lactate and pediatric sepsis. Am J Emerg Med 2020; 38:2620.
  87. Hunter CL, Silvestri S, Ralls G, et al. A prehospital screening tool utilizing end-tidal carbon dioxide predicts sepsis and severe sepsis. Am J Emerg Med 2016; 34:813.
  88. Hunter CL, Silvestri S, Ralls G, et al. Comparing Quick Sequential Organ Failure Assessment Scores to End-tidal Carbon Dioxide as Mortality Predictors in Prehospital Patients with Suspected Sepsis. West J Emerg Med 2018; 19:446.
Topic 273 Version 27.0