Characteristics of commonly used filters for salvaged autologous red blood cells

Characteristics of commonly used filters for salvaged autologous red blood cells
  Haemonetics pall SQ40S microaggregate filter Haemonetics RS leukocyte reduction filter Haemonetics lipiguard transfusion filter for salvaged blood
Cost $ $$ $$
Capacity NA, but demonstrates greater capacity than the other 2 filters 1 unit of washed or unwashed salvaged blood 2 bowls or 450 mL
Hold up volume 20 mL 35 mL 20 mL
Screen filtrations size 40 microns 40 microns 40 microns
Leukocyte removal NA ~99% ~71%
Fat globule removal NA ~84% ~84%
C3a reduction NA NA ~40%
Flow rate ~500 mL/minute at 150 mmHg About 80 mL/minute at 300 mmHg About 80 mL/minute at 300 mmHg
Graphic 143664 Version 1.0