Apnea-CO2 alarm

Apnea-CO2 alarm
Manifestations
  • Loss of capnogram
Potential causes
  • Patient factors
    • Patient apnea (no respiratory effort)
    • Esophageal intubation
    • Cardiac arrest
  • Equipment factors
    • False alarm
      • Capnometer sample line crack or disconnect
    • Exhaled gas exiting breathing system before sampling site
      • Large cuff leak (no exhaled gas into breathing system)
      • Lower airway suction (gastric tube in trachea, tracheal/bronchial suction)
    • Breathing system disconnect
    • Breathing system obstruction
    • Airway (ETT or LMA) obstruction
    • Ventilator turned off or cycling failure malfunction
Suggested responses
  • Check circulation (electrocardiogram and plethysmogram)
  • Determine whether patient ventilation is occurring (examine chest, auscultate lungs, check respiratory monitors, check SpO2)
  • Rule out esophageal intubation if recently intubated
  • Look for breathing circuit disconnection or obstruction
  • Verify that ventilator is operating
  • Switch to manual ventilation
  • If necessary, ventilate with manual resuscitation device connected to O2
CO2: carbon dioxide; ETT: endotracheal tube; LMA: laryngeal mask airway; SpO2: pulse oximetry; O2: oxygen.