Apnea alarm overview

Apnea alarm overview
Mechanical ventilation
  • Apnea - pressure only
    • False alarm – threshold set too high
    • Breathing system leak
    • Pressure sensor tube disconnect or obstruction
  • Apnea - flow only
    • False alarm – high threshold (eg, neonate)
    • Malfunctioning flow sensor (eg, due to moisture or broken filament)
  • Apnea-CO2 only
    • Capnometer sample line crack or disconnect
    • Esophageal intubation
    • Cardiac arrest
  • Apnea-flow and apnea-CO2
    • Breathing system obstruction
    • Large cuff leak (no exhaled gas into breathing system)
    • Esophageal intubation
  • Apnea-pressure and apnea-flow and apnea-CO2
    • Breathing system disconnect
    • Ventilator turned off or cycling failure malfunction
    • Lower airway suction (gastric tube in trachea, tracheal/bronchial suction)
Spontaneous or manual ventilation
  • Apnea-pressure only
    • False alarm – When switching from mechanical to spontaneous ventilation (apnea-pressure not applicable in spontaneous mode)
  • Apnea-flow only
    • False alarm – high threshold (neonate)
    • Incomplete endotracheal tube kink or obstruction
    • Breathing system disconnect with patient still connected to CO2 sampling site
    • Malfunctioning flow sensor (eg, due to moisture or broken filament)
  • Apnea-CO2 only
    • Capnometer sample line crack or disconnect
  • Apnea-flow and apnea-CO2
    • Patient apnea
    • Breathing system obstruction
    • Airway (ETT or LMA) obstruction
    • Breathing system disconnect
    • Lower airway suction (gastric tube in trachea, tracheal/bronchial suction)
    • Esophageal intubation
    • Cardiac arrest
Differential diagnosis based on mode of ventilation and combination of alarms.
CO2: carbon dioxide; ETT: endotracheal tube; LMA: laryngeal mask airway.